Before Use

Inspect the transducer for cracks and cuts. Perform Conductivity and Leakage Testing in the following order: Conductivity test, Bite Guard, and Leakage test.

If Conductivity fails, do not perform the leakage test. If leakage test fails, take out of circulation and call Service Representative.

Storage

Do NOT store the transducer while wet. Do NOT store in extreme temperatures or direct sunlight.

Use a Siemens Approved disposable tip protector. Store in clean, dry, and protected environment.

Transducer should NOT be stored in transducer holders on the system.