Design Change and Recommendations
Medtronic has implemented a design change to correct the issue of Model 37751 Rechargers becoming unresponsive and beeping. Devices with the design change are now included in recharging systems for Spinal Cord Stimulation (SCS) and Deep Brain Stimulation (DBS).
For products manufactured prior to the design change: To prevent the unresponsive error state, plug the Recharger into the AC power supply before starting a recharging session and remain connected until the session is finished.
For products manufactured after the design change: Follow the instructions included with the Charging System user manual shipped with the device.
Important Safety Information
If the neurostimulator battery is allowed to become fully depleted, it can lead to loss of therapy and return of associated disease-specific symptoms.