Introduction to the Paradigm Insulin Infusion Pump
Important Safety Information
Avoid immersing your pump in water, as it can be damaged. The pump was tested to meet requirements for IPX7 at the time of manufacture.
Be aware that water, lotions, sunscreens, and insect repellent can damage the pump case.
Electrostatic Discharge
Although the pump is designed to be unaffected by typical levels of electrostatic discharge, very high levels can reset the pump’s software, causing error alarms.
If you experience an error alarm, press the ESC and ACT buttons to clear it. If unable, remove and replace the battery. After clearing, verify date, time, and settings are correct, as the software reset may have erased previous settings.
Please report any errors or problems to the 24-Hour HelpLine at 1-800-MINIMED (1-800-646-4633).