Field Safety Notice: High Battery Impedance May Initiate Safety Mode

Boston Scientific has determined that dual chamber INGENIOTM family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life when the device’s battery exhibits high internal impedance.

This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing.

Affected Devices

Approximately 48,000 active dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population.

No affected devices remain available for implant.

Recommended Actions

If a device enters Safety Mode, schedule replacement. In situations where non-programmable Safety Mode pacing parameters may not provide optimal support of patient’s cardiac condition, consider early device replacement.