Safety Precautions
Observe precautions to ensure safety of patients and operators during cleaning, disinfection, and sterilization of the transducer.
Never immerse the transducer connector or non-waterproof sections into liquids to avoid electric shock.
Precautions Concerning Cleaning, Disinfection, and Sterilization
Wear protective gloves during cleaning, disinfection, and sterilization to prevent infection.
Rinse the transducer thoroughly with purified water after cleaning and with sterile or deionized water after disinfection to remove chemical residues.
Dry the transducer surface thoroughly after chemical cleaning or disinfection.
Confirm the effectiveness of each chemical for cleaning, disinfection, or sterilization based on the manufacturer’s documentation.
Chemical Hazard (for USA only)
Observe instructions to protect patients and operators from inflammation or poisoning by chemical substances.
Precautions Concerning Restriction of Sale or Use (for USA only)
United States law restricts this device to sale or use by, or on the order of a physician.