Welcome to SmartGuard Auto Mode

This guide introduces SmartGuard technology, which automatically adjusts basal insulin delivery based on sensor glucose (SG) values.

In Auto Mode, basal insulin is delivered based on SG values and recent insulin delivery needs, with a target of 120 mg/dL.

You can temporarily change the target to 150 mg/dL for exercise or other times.

Auto Mode may recommend a correction bolus depending on insulin needs, and you will receive a BG required alert if your pump needs a BG to enter or stay in Auto Mode.

Important Notes

Manual Mode is not a mode that you turn on or off in a menu, but is simply the mode the pump is in when it is not in Auto Mode.

There are times in Auto Mode when basal insulin is being delivered according to recent insulin needs, but is not being adjusted based on an SG reading, called Safe Basal.

Manual: