Corrective and Preventive Actions (CAPA)

The ED-S3OXT Duodenoscope has several issues related to corrective and preventive actions, including:

  • Lack of investigations into failed, returned, and repaired duodenoscopes.
  • Inadequate procedures for receiving, reviewing, and evaluating complaints.
  • Lack of root-cause investigations into returned duodenoscopes with recurring issues.

Complaints and Corrective Actions

Several complaints have been reported, including:

  • Stuck suction buttons.
  • Incompatibility with EPX-2500 Processor.
  • Water leaks and image failures.

Validation of Reprocessing Instructions

The firm’s reprocessing instructions for the ED-S3OXT Duodenoscope have not been adequately validated, leading to concerns about the effectiveness of the cleaning process.