Important Safety Notification
ResMed has issued a field safety notification regarding the Astral ventilator’s internal battery, which may experience false fuse activation, leading to power loss.
The issue may cause the device to shut down, but ventilation will continue if the device is connected to external power. If not, ventilation will stop with the activation of the total power fail alarm.
Affected Products
Astral 100, Astral 100SC, and Astral 150 ventilators, as well as Astral External Batteries and battery packs (spares), are affected.
Actions Required
Users should continue to follow the precautions detailed in the user manual, ensure alternate ventilation equipment is available, and return devices for routine maintenance and battery replacement.